Multidimensional fetal flow imaging with cardiovascular magnetic resonance: a feasibility study

J Cardiovasc Magn Reson. 2018 Nov 29;20(1):77. doi: 10.1186/s12968-018-0498-z.

Abstract

Purpose: To image multidimensional flow in fetuses using golden-angle radial phase contrast cardiovascular magnetic resonance (PC-CMR) with motion correction and retrospective gating.

Methods: A novel PC-CMR method was developed using an ungated golden-angle radial acquisition with continuously incremented velocity encoding. Healthy subjects (n = 5, 27 ± 3 years, males) and pregnant females (n = 5, 34 ± 2 weeks gestation) were imaged at 3 T using the proposed sequence. Real-time reconstructions were first performed for retrospective motion correction and cardiac gating (using metric optimized gating, MOG). CINE reconstructions of multidimensional flow were then performed using the corrected and gated data.

Results: In adults, flows obtained using the proposed method agreed strongly with those obtained using a conventionally gated Cartesian acquisition. Across the five adults, bias and limits of agreement were - 1.0 cm/s and [- 5.1, 3.2] cm/s for mean velocities and - 1.1 cm/s and [- 6.5, 4.3] cm/s for peak velocities. Temporal correlation between corresponding waveforms was also high (R~ 0.98). Calculated timing errors between MOG and pulse-gating RR intervals were low (~ 20 ms). First insights into multidimensional fetal blood flows were achieved. Inter-subject consistency in fetal descending aortic flows (n = 3) was strong with an average velocity of 27.1 ± 0.4 cm/s, peak systolic velocity of 70.0 ± 1.8 cm/s and an intra-class correlation coefficient of 0.95 between the velocity waveforms. In one fetal case, high flow waveform reproducibility was demonstrated in the ascending aorta (R = 0.97) and main pulmonary artery (R = 0.99).

Conclusion: Multidimensional PC-CMR of fetal flow was developed and validated, incorporating retrospective motion compensation and cardiac gating. Using this method, the first quantification and visualization of multidimensional fetal blood flow was achieved using CMR.

Keywords: Compressed sensing; Fetal; Golden angle radial; Motion correction; Phase contrast MRI; Retrospective gating.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study
  • Video-Audio Media

MeSH terms

  • Adult
  • Aorta / diagnostic imaging
  • Aorta / physiopathology
  • Blood Flow Velocity
  • Cardiac-Gated Imaging Techniques
  • Case-Control Studies
  • Coronary Circulation*
  • Feasibility Studies
  • Female
  • Fetal Heart / diagnostic imaging*
  • Fetal Heart / physiopathology
  • Gestational Age
  • Humans
  • Image Interpretation, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Myocardial Perfusion Imaging / methods*
  • Predictive Value of Tests
  • Pregnancy
  • Prenatal Diagnosis / methods*
  • Pulmonary Artery / diagnostic imaging
  • Pulmonary Artery / physiopathology
  • Reproducibility of Results

Grants and funding