Possible artifacts in dynamic CEST MRI due to motion and field alterations

J Magn Reson. 2019 Jan:298:16-22. doi: 10.1016/j.jmr.2018.11.002. Epub 2018 Nov 13.

Abstract

Purpose: Dynamic CEST studies such as dynamic glucose enhanced imaging, have gained a lot of attention recently. The expected CEST effects after injection are rather small in tissue especially at clinical field strengths (0.5-2%). Small movements during the dynamic CEST measurement together with a subtraction-based evaluation can lead to pseudo CEST effects of the same order of magnitude. These artifacts are studied herein.

Methods: A brain tumor patient 3D-CEST baseline scan without glucose injection performed at 3 T is used to generate a virtual dynamic measurement introducing different kinds of simulated motion and B0 shifts.

Results: Minor motion (0.6 mm translations) and B0 artifacts (7 Hz shift) can lead to pseudo effects in the order of 1% in dynamic CEST imaging. Especially around tissue interfaces such as CSF borders or tumor affected areas, the pseudo effect patterns are non-intuitive and can be mistaken as dynamic agent uptake.

Conclusion: Correction or mitigation for small motions is crucial for dynamic CEST imaging, especially in subjects with lesions. Concomitant B0 alterations can as well induce pseudo CEST effects at 3 T.