Purpose: We aim to compare the results of spin echo-echo planar imaging (SE-EPI)-based T2 mapping with those of the conventional Carr-Purcell-Meiboom-Gill (CPMG) method and to investigate the potential validity of SE-EPI-T2 mapping for the characterization of prostate cancer (PCa).
Methods: Our retrospective study included 42 PCa patients and 42 noncancer patients who underwent 3.0T MRI with b values ranging from 0 to 2000 s/mm2 and echo times (TEs) ranging from 32 to 100 ms before biopsies. Bland-Altman analysis was used to compare the agreement between the two methods. The correlations between CPMG-T2 values and SE-EPI-T2 values at different b values were determined by Spearman's rho analysis or Pearson analysis. The Mann-Whitney U test and two-sample t tests were used to analyze the differences between the cancerous and noncancerous groups.
Results: Substantial agreement regarding the measurements was observed between the two methods. The average correlation between the CPMG-T2 values and SE-EPI-T2 values was moderate and positive, and the best correlations were found at b = 200 s/mm2 in the noncancer group (r = 0.557, P = 0.000) and at b = 100 s/mm2 in the cancer group (r = 0.537, P = 0.000). In addition, statistically significant differences were found between the noncancer and cancer groups in T2 values and ADC values (diff TEs) (P = 0.000).
Conclusions: Substantial agreement in the measurements was found between the SE-EPI method and CPMG method. SE-EPI-based T2 mapping has potential clinical value for the prostate and can be considered an alternative to the traditional CPMG-T2 mapping method.
Keywords: Diffusion magnetic resonance imaging; Echo planar imaging; Magnetic resonance imaging; Prostate cancer; T2 mapping.