Double stent assisted coiling of intracranial bifurcation aneurysms in Y and X configurations with the Neuroform ATLAS stent: immediate and mid term angiographic and clinical follow-up

J Neurointerv Surg. 2019 Dec;11(12):1239-1242. doi: 10.1136/neurintsurg-2019-015175. Epub 2019 Jul 27.

Abstract

Purpose: Self-expandable stents have broadened the spectrum of endovascular treatment of intracranial aneurysms. However, procedures involving double stenting in Y/X configurations carry a relatively high risk of procedural complications. The Neuroform ATLAS, the evolution of Neuroform EZ, is a nitinol self-expanding hybrid/open cell stent which can be delivered through a low profile 0.017 inch catheter. We present our experience in the treatment of intracranial aneurysms with this stent in Y and X configurations.

Materials and methods: We prospectively maintained a database from consecutive patients who underwent double stent assisted coiling with the Neuroform ATLAS, from July 2015 to February 2019. Clinical and angiographic results were analyzed.

Results: 55 patients harboring 55 bifurcation aneurysms were treated with double stenting: 52 'Y' configurations, 3 'X' configurations. Deployment was successful in all cases. Post-treatment control angiography showed complete occlusion in 33 cases (60%), neck remnant in 8 cases (14.5%), and incomplete occlusion in 14 cases (25.4%). The overall symptomatic periprocedural complication rate was 12.7%. 38 aneurysms underwent follow-up (69%, mean duration 16 months): 33 aneurysms (87%) were completely occluded, 3 aneurysms (8%) had a neck remnant, and 2 aneurysms (5%) were incompletely occluded.

Conclusion: The Neuroform ATLAS is an effective device for treatment of bifurcation aneurysms, allowing good conformability, a high level of navigability, and easy mesh crossing to perform Y/X stenting procedures. The rate of procedural complications remains non-negligible, and an indication for a double stenting procedure should be carefully discussed in a multidisciplinary meeting.

Keywords: aneurysm.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Cerebral Angiography / instrumentation
  • Cerebral Angiography / methods*
  • Embolization, Therapeutic / methods
  • Female
  • Follow-Up Studies
  • Humans
  • Intracranial Aneurysm / diagnostic imaging*
  • Intracranial Aneurysm / surgery*
  • Male
  • Middle Aged
  • Prospective Studies
  • Retrospective Studies
  • Self Expandable Metallic Stents*
  • Treatment Outcome