Objective: In implantable cardiac defibrillators (ICDs), long-detection times safely reduce unnecessary and inappropriate therapies. We aimed to evaluate ICD treatment of ventricular arrhythmias in women, compared with men, also taking into account ICD detection.
Methods: The Advance III trial randomised patients implanted with an ICD for primary or secondary prevention in two arms-long and nominal ventricular arrhythmias detection times before therapy delivering (number of intervals needed to detect (NID) 30/40 and 18/24, respectively). The main endpoint of this post hoc analysis was the incidence of ICD therapies evaluated through Kaplan-Meier method and univariate Cox regression models.
Results: Overall, 1902 patients (304 women, 65±11 years) were randomised. Women showed a lower risk of ICD therapy (HR 0.63, 95% CI 0.43 to 0.93, p=0.022); this difference was observed only in the long-detection arm (HR 0.37, p=0.013) and not in the short detection arm (HR 0.82, p=0.414). No significant sex differences were observed concerning inappropriate therapies and mortality rate. Long-detection settings significantly reduced overall ICD therapies and appropriate ICD therapies, both in women (overall HR 0.31, p=0.007; appropriate HR 0.33, p=0.033) and in men (overall HR 0.69, p=0.006; appropriate HR 0.73, p=0.048).
Conclusions: In patients with ICDs, the strategy of setting a long-detection time to treat ventricular arrhythmias (NID 30/40) reduces overall delivered therapies, both in women and men, when compared with nominal setting (NID 18/24). The reduction was significantly higher in women. Overall, women were less likely to experience ICD therapies than men; this result was only observed in the long-detection arm.
Clinical trial registration: NCT00617175.
Keywords: heart failure with reduced ejection fraction; implanted cardiac defibrillators; ventricular fibrillation; ventricular tachycardia.