Aim: To assess the fat content of the pancreas using quantitative computed tomography (QCT) and to correlate the results with chemical-shift-encoded magnetic resonance imaging (CSE-MRI) measurements of proton density fat fraction (PDFF).
Material and methods: Institutional review board approval for this research was obtained and 52 participants (25 men, 27 women; mean age 35.1 years; age range 22-50 years), who were enrolled in the Prospective Urban Rural Epidemiology (PURE) Study, underwent QCT and CSE-MRI for quantification of fat content in the pancreas. Two observers placed regions of interest (area of 100-130 mm2) in the head, body, and tail of the pancreas as closely matched as possible on the two scans. Pearson correlation and Bland-Altman analysis were performed to evaluate the correlation between the QCT and CSE-MRI measurements and the systematic difference between the two techniques.
Results: The QCT and CSE-MRI measurements of pancreatic fat content were well correlated (r=0.805, p<0.0001), although Bland-Altman analysis showed that the QCT measurements were systematically lower by 6.3% compared to CSE-MRI PDFF.
Conclusion: In conclusion, the results of this study suggest good correlation between QCT and CSE-MRI measurements of pancreatic fat content. Further studies are required to improve the numerical agreement of QCT measurements with PDFF.
Copyright © 2020 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.