Validation of a New Method for 2D Fusion Imaging Registration in a System Prepared Only for 3D

J Endovasc Ther. 2020 Jun;27(3):468-472. doi: 10.1177/1526602820912223. Epub 2020 Mar 20.

Abstract

Purpose: To validate a new 2D-3D registration method of fusion imaging during aortic repair in a system prepared only for 3D-3D registration and to compare radiation doses and accuracy. Materials and Methods: The study involved 189 patients, including 94 patients (median age 70 years; 85 men) who underwent abdominal endovascular aneurysm repair (EVAR) with 2D-3D fusion on an Artis zee imaging system and 95 EVAR patients (median age 70 years; 81 men) from a prior study who had 3D-3D registration done using cone beam computed tomography (CBCT). For the 2D-3D registration, an offline CBCT of the empty operating table was imported into the intraoperative dataset and superimposed on the preoperative computed tomography angiogram (CTA). Then 2 intraoperative single-frame 2D images of the skeleton were aligned with the patient's skeleton on the preoperative CTA to complete the registration process. A digital subtraction angiogram was done to correct any misalignment of the aortic CTA volume. Values are given as the median [interquartile range (IQR) Q1, Q3]. Results: The 2D-3D registration had an accuracy of 4.0 mm (IQR 3.0, 5.0) after bone matching compared with the final correction with DSA (78% within 5 mm). By applying the 2D-3D protocol the radiation exposure (dose area product) from the registration of the fusion image was significantly reduced compared with the 3D-3D registration [1.12 Gy∙cm2 (IQR 0.41, 2.14) vs 43.4 Gy∙cm2 (IQR 37.1, 49.0), respectively; p<0.001). Conclusion: The new 2D-3D registration protocol based on 2 single-frame images avoids an intraoperative CBCT and can be used for fusion imaging registration in a system originally designed for 3D-3D only. This 2D-3D registration protocol is accurate and leads to a significant reduction in radiation exposure.

Keywords: computed tomography angiography; cone beam computed tomography; endovascular aneurysm repair; fusion imaging; procedure guidance; radiation exposure; registration protocol.

Publication types

  • Comparative Study
  • Validation Study

MeSH terms

  • Aged
  • Anatomic Landmarks
  • Aorta, Abdominal / diagnostic imaging*
  • Aorta, Abdominal / surgery
  • Aortic Aneurysm, Abdominal / diagnostic imaging*
  • Aortic Aneurysm, Abdominal / surgery
  • Aortography* / adverse effects
  • Blood Vessel Prosthesis Implantation
  • Computed Tomography Angiography* / adverse effects
  • Endovascular Procedures
  • Female
  • Humans
  • Imaging, Three-Dimensional* / adverse effects
  • Male
  • Predictive Value of Tests
  • Radiation Dosage
  • Radiation Exposure / adverse effects
  • Radiation Exposure / prevention & control
  • Reproducibility of Results
  • Risk Factors