Purpose: To investigate the diagnostic accuracy of applying four levels of manual pressure in Shear Wave Elastography (SWE) of the breast and to assess inter-rater reliability.
Materials and methods: Single-center prospective preliminary study including patients receiving ultrasound examination of breast lesions as part of routine clinical practice. SWE was performed on 60 breast masses (26 benign and 34 malignant) in 54 patients by a breast fellowship trained radiologist. Stiffness values were compared between benign and malignant masses at four levels of manual compression: none, mild, moderate, and marked. Accuracy of SWE was assessed using receiving operating characteristics analysis at each level. In 18 patients, a second radiologist repeated the SWE acquisitions to evaluate reproducibility. Reproducibility was assessed using intraclass correlation coefficient.
Results: Without compression, we observed no significant difference in stiffness (p > 0.99) between benign and malignant lesions, and SWE demonstrated low accuracy (area under the curve = 0.64). Stiffness was higher in malignant lesions at all levels of compression (p < 0.001). SWE demonstrated good accuracy at all three levels of compression (from area under the curve = 0.71 to 0.84 across Emax and Emean), with high interobserver agreement.
Conclusion: This preliminary study suggests that not using compression during SWE for breast lesion characterization offers suboptimal results. On the contrary, application of compression yields high diagnostic performance with good interobserver agreement and, as such, should be included in routine clinical practice.
Keywords: Breast lesion characterization; Compression; Shear wave elastography; Variability.
Copyright © 2020 The Association of University Radiologists. All rights reserved.