False Elevation of Volume Determined by Bladder Scanner Secondary to Bowel Obstruction

Clin Pract Cases Emerg Med. 2020 Feb 24;4(2):158-160. doi: 10.5811/cpcem.2019.12.45103. eCollection 2020 May.

Abstract

Bladder scanners allow for quick determination of bladder volumes (BV) with minimal training. BV measured by a machine is generally accurate; however, circumstances exist in which falsely elevated BVs are reported. This case details a patient with a significant small bowel obstruction (SBO) due to superior mesenteric artery syndrome causing a falsely elevated BV. We believe this is the first case report of a SBO causing an elevated BV by bladder scanner. Emergency physicians should be aware of the pitfalls of using bladder scanners, and use their point-of-care ultrasound skills when possible to expand their differential.

Publication types

  • Case Reports