Motor unit number index as an individual biomarker: Reference limits of intra-individual variability over time in healthy subjects

Clin Neurophysiol. 2020 Sep;131(9):2209-2215. doi: 10.1016/j.clinph.2020.06.019. Epub 2020 Jul 9.

Abstract

Objective: Motor unit number index (MUNIX) is proposed to monitor neuromuscular disorders. Our objective is to determine the intra-individual variability over time of the MUNIX.

Methods: In 11 different hospital centres, MUNIX was assessed twice, at least 3 months apart (range 90-360 days), in tibialis anterior (TA), abductor pollicis brevis (APB), abductor digiti minimi (ADM) and deltoid muscles in 118 healthy subjects. MUNIX sum score 2, 3 and 4 were respectively the sum of the MUNIX of the TA and ADM, of the TA, APB and ADM and of the TA, APB, ADM and deltoid muscles.

Results: The repeatability of the MUNIX was better for sum scores than for single muscle recordings. The variability of the MUNIX was independent of sex, age, interval between measurements and was lower for experienced than non-experienced operators. The 95th percentile of the coefficient of variability of the MUNIX sum score 2, 3 and 4 were respectively 22%, 18% and 15% for experienced operators.

Conclusions: The MUNIX technique must be performed by experienced operators on several muscles to reduce its variability and improve its reliability.

Significance: A variation of the MUNIX sum score ≥20% can be interpreted as a significant change of muscle innervation.

Keywords: Healthy subjects; MUNIX; Normal values; Repeatability; Variability.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Electromyography / methods
  • Female
  • Healthy Volunteers
  • Humans
  • Male
  • Middle Aged
  • Motor Neurons / physiology*
  • Muscle, Skeletal / physiology*
  • Neuromuscular Diseases / diagnosis
  • Neuromuscular Diseases / physiopathology
  • Recruitment, Neurophysiological / physiology*
  • Reproducibility of Results
  • Young Adult