A Polyvinyl Alcohol-Based Thermochromic Material for Ultrasound Therapy Phantoms

Ultrasound Med Biol. 2020 Nov;46(11):3135-3144. doi: 10.1016/j.ultrasmedbio.2020.07.032. Epub 2020 Aug 29.

Abstract

Temperature estimation is a fundamental step in assessment of the efficacy of thermal therapy. A thermochromic material sensitive within the temperature range 52.5°C-75°C has been developed. The material is based on polyvinyl alcohol cryogel with the addition of a commercial thermochromic ink. It is simple to manufacture, low cost, non-toxic and versatile. The thermal response of the material was evaluated using multiple methods, including immersion in a temperature-controlled water bath, a temperature-controlled heated needle and high-intensity focused ultrasound (HIFU) sonication. Changes in colour were evaluated using both RGB (red, green, blue) maps and pixel intensities. Acoustic and thermal properties of the material were measured. Thermo-acoustic simulations were run with an open-source software, and results were compared with the HIFU experiments, showing good agreement. The material has good potential for the development of ultrasound therapy phantoms.

Keywords: High-intensity focused ultrasound; Phantom; Quality assurance; Thermal ablation; Thermochromic material.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Color
  • Hot Temperature
  • Materials Testing*
  • Phantoms, Imaging*
  • Polyvinyl Alcohol*
  • Ultrasonic Therapy*

Substances

  • Polyvinyl Alcohol