A prototype Multi-X-ray-source array (MXA) for digital breast tomosynthesis

Phys Med Biol. 2020 Dec 18;65(23):235033. doi: 10.1088/1361-6560/abc305.

Abstract

The design and testing of a prototype Multi-X-ray-source Array (MXA) for digital breast tomosynthesis is reported. The MXA is comprised of an array of tungsten filament cathodes with focus cup grid-controlled modulation and a common rotating anode housed in a single vacuum envelope. Prototypes consisting of arrays of three-source elements and eleven-source-elements were fabricated and evaluated. The prototype sources demonstrated focal spot sizes of 0.3 mm at 45 kV with 50 mA. Measured x-ray spectra were consistent with the molybdenum anode employed, and the tube output (air kerma) was between 0.6 mGy/100 mAs at 20 kV and 17 mGy/100 mAs at 45 kV with a distance of 100 cm. HVL measurements ranged from 0.5 mm Al at 30 kV to 0.8 mm Al at 45 kV, and x-ray pulse widths were varied from 20 ms to 110 ms at operating frequencies ultimately to be limited by source turn-on/off times of ∼1 ms. Initial results of reconstructed tomographic data are presented.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Breast Neoplasms / diagnostic imaging*
  • Female
  • Humans
  • Mammography / instrumentation*
  • Mammography / methods*
  • Phantoms, Imaging*
  • Tomography, X-Ray Computed / instrumentation*
  • Tomography, X-Ray Computed / methods*