ClearSight™ finger cuff versus invasive arterial pressure measurement in patients with body mass index above 45 kg/m2

BMC Anesthesiol. 2021 May 18;21(1):152. doi: 10.1186/s12871-021-01374-x.

Abstract

Background: Measuring blood pressure in patients with obesity is challenging. The ClearSight™ finger cuff (FC) uses the vascular unloading technique to provide continuous non-invasive blood pressure measurements. We aimed to test the agreement of the FC with invasive radial arterial monitoring (INV) in patients with obesity.

Methods: Participants had a body mass index (BMI) ≥45 kg/m2 and underwent laparoscopic bariatric surgery. FC and INV measurements were obtained simultaneously every 5 min on each patient, following induction of anesthesia. Agreement over time was assessed using modified Bland-Altman plots and error grid analysis permitted clinical interpretation of the results. Four-quadrant plots allowed assessment of concordance in blood pressure changes.

Results: The 30 participants had a median (IQR) BMI of 50.2 kg/m2 (IQR 48.3-55.3). The observed bias (SD, 95% limits of agreement) for systolic blood pressure (SBP) was 14.3 mmHg (14.1, -13.4 - 42.0), 5.2 mmHg (10.9, -16.0 - 26.5) for mean arterial pressure (MAP) and 2.6 mmHg (10.8, -18.6 - 23.8) for diastolic blood pressure (DBP). Error grid analysis showed that the proportion of readings in risk zones A-E were 90.8, 6.5, 2.7, 0 and 0% for SBP and 91.4, 4.3, 4.3, 0 and 0% for MAP, respectively. Discordance occurred in ≤8% of pairs for consecutive change in SBP, MAP and DBP.

Conclusions: The vascular unloading technique was not adequately in agreement with radial arterial monitoring. Evaluation in a larger sample is required before recommending this technique for intraoperative monitoring of patients with BMI ≥45 kg/m2.

Keywords: ClearSight™; Invasive blood pressure; Non-invasive blood pressure; Obesity; Vascular unloading.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Bariatric Surgery
  • Blood Pressure / physiology
  • Blood Pressure Determination / instrumentation*
  • Blood Pressure Determination / methods*
  • Body Mass Index
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Intraoperative / instrumentation*
  • Monitoring, Intraoperative / methods*
  • Obesity / physiopathology
  • Obesity / surgery*
  • Prospective Studies
  • Radial Artery / physiology*
  • Reproducibility of Results
  • Young Adult