Impact of 24/7 loading of blood culture bottles in a new automated incubator on the diagnosis of bloodstream infections

Eur J Clin Microbiol Infect Dis. 2021 Dec;40(12):2639-2643. doi: 10.1007/s10096-021-04283-6. Epub 2021 Jun 1.

Abstract

Blood culturing (BC) remains the gold standard for bloodstream diagnosis but its workflow is slow. We aimed reducing this time by implementing a new automated incubator with a 24/7 BC workflow. With this new strategy, time to incubation was shorter (1.52 h vs 6.82 h), positivity rates were higher (10.6% vs 8.9%, p<0.05), and the number of BSI diagnostics increased (16.1% vs 13.8% patients and 2.3 vs 1.9 density episode per 1000 hospital days). Our results show that implementing automatic loading of BC bottles with a 24/7 strategy not only shortened time to diagnosis but significantly increased the BSI diagnosis rate.

Keywords: 24/7; Blood culture; Bloodstream infection; diagnosis.

Publication types

  • Evaluation Study

MeSH terms

  • Automation / instrumentation
  • Automation / methods*
  • Bacteremia / diagnosis*
  • Bacteremia / microbiology*
  • Bacteria / growth & development*
  • Bacteria / isolation & purification
  • Blood Culture / instrumentation
  • Blood Culture / methods*
  • Humans
  • Incubators
  • Time Factors