Objectives: To characterize hemodynamics of serial coronary stenoses using fractional flow reserve (FFR) pullbacks and the pullback pressure gradients (PPG) index.
Background: The cross-talk between stenoses within the same coronary artery makes the prediction of the functional contribution of each lesion challenging.
Methods and results: One-hundred seventeen patients undergoing coronary angiography for stable angina were prospectively recruited. Serial lesions were defined as two or more narrowings with visual diameter stenosis >50% on conventional angiography. Motorized FFR pullback tracings were obtained at 1 mm/s. Pullbacks were visually adjudicated as presenting two, one, and no focal pressure drops. The pattern of disease (i.e., focal or diffuse) was quantified using the PPG index. Twenty-five vessels presented serial lesions (mean PPG 0.48 ± 0.17). Two, one or no focal pressure drops were observed in 40% (n = 10; PPG 0.59 ± 0.17), 52% (n = 13; PPG 0.44 ± 0.12) and 8% of cases (n = 2; PPG 0.27 ± 0.01; p-value = 0.01). Distal FFR was similar between vessels with two, one and no focal pressure drops in the pullback curve (p-value = 0.27). The PPG index independently predicted the presence of two focal pressure drops in the pullback curve (p = 0.04).
Conclusions: FFR pullbacks in serial coronary lesions exhibit three distinct functional patterns. High PPG was associated with pullback curves presenting two pressure drops. The PPG provides a quantitative assessment of the pattern of coronary artery disease in cases with serial lesions and might be useful to assess the appropriateness of percutaneous revascularization.
Keywords: coronary artery disease; coronary physiology; fractional flow reserve; pressure drops.