Spatial frequency domain imager based on a compact multiaperture camera: testing and feasibility for noninvasive burn severity assessment

J Biomed Opt. 2021 Aug;26(8):086001. doi: 10.1117/1.JBO.26.8.086001.

Abstract

Significance: Spatial frequency domain imaging (SFDI) is a wide-field imaging technique that provides quantitative maps of tissue optical properties. We describe a compact SFDI imager that employs a multispectral compound-eye camera. This design enables simultaneous image acquisition at multiple wavelengths. Such a device has potential for application for quantitative evaluation of superficial tissues by nonspecialists in low-resource settings.

Aim: The aim of this work was to develop a compact SFDI imager for widefield imaging of in-vivo tissue optical properties and verify its ability to measure optical properties of tissue-simulating phantoms and in a preclinical model of burn wounds.

Approach: This compound-eye imager was constructed using a CMOS sensor subdivided into multiple regions, each having a bandpass filter and objective lens. The ability of the instrument to image optical properties was compared with (1) a commercial SFDI imager and (2) a laboratory-based system. Initial validation of ability to accurately characterize optical properties was performed using a tissue-simulating optical phantom. It was then applied to an established murine model of thermal contact burn severity. In-vivo measurements of the optical properties of rat skin were performed before and after the application of burns. Histology was used to verify burn severity.

Results: Measurements of the tissue-simulating phantom optical properties made using the compound-eye imager agree with measurements made using the two comparison SFDI devices. For the murine burn model, the burns showed a decrease in the reduced scattering coefficient at all measurement wavelengths compared with preburn measurements at the same locations. This is consistent with previously reported changes in scattering that occur in full-thickness burns.

Conclusion: We demonstrate the potential for SFDI to be translated into compact form factor using a compound-eye camera that is capable of obtaining multiple wavelengths channels simultaneously.

Keywords: burn wounds; compound eye; diffuse optics; multispectral imaging; spatial frequency domain imaging; thin observational module by bound optics.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Burns* / diagnostic imaging
  • Feasibility Studies
  • Mice
  • Optical Imaging*
  • Phantoms, Imaging
  • Rats
  • Skin / diagnostic imaging