Transient out-of-range impedance in "hybrid" implantable cardioverter-defibrilator system: A case series

Pacing Clin Electrophysiol. 2022 Feb;45(2):270-273. doi: 10.1111/pace.14401. Epub 2021 Dec 10.

Abstract

A retrospective analysis of 60 patients with hybrid implantable cardioverter-defibrilator (ICD) systems: Boston Scientific device paired with non-Boston leads. In 10 (17%) patients transient, out-of-range peaks of ventricular pace impedance trend were observed. Probable cause is header-lead interaction incompatibility. This matter is known mainly for pacemakers systems but not for ICDs. Investigation this issue is crucial because consequences in ICD systems are unpredictable and risk might be higher than in pacing systems.

Keywords: ICD; Lead failure; header-lead interaction; hybrid CIED system; pacing impedance.

MeSH terms

  • Aged
  • Defibrillators, Implantable*
  • Diagnosis, Differential
  • Electric Impedance
  • Equipment Failure Analysis*
  • Female
  • Humans
  • Male
  • Prosthesis Design
  • Retrospective Studies