Three internal fixation methods for Danis-Weber-B distal fibular fractures: A biomechanical comparison in an osteoporotic fibula model

Foot Ankle Surg. 2022 Oct;28(7):845-851. doi: 10.1016/j.fas.2021.11.006. Epub 2021 Nov 18.

Abstract

A common agreement for the surgical treatment of osteoporotic ankle fractures has not been defined yet although locking plates are preferred for fractures with poor bone quality. This study aims to evaluate the mechanical stability of locked and conventional plates on osteoporotic Danis-Weber-B-fibula fracture models. Fractured custom-made osteoporotic fibulae were treated with neutralization plate plus lag screw, locking plate plus lag screw, or a standalone locking plate. Load until failure was applied mimicking single-leg stance. Stiffness, failureload, and interfragmentary movements were investigated. Stiffness, failureload and axial fragment movement showed no significant differences among groups. Shear movements and fragment rotation around the shaft of the neutralization plate were on average twice as high as those of the locking plates. Although no superiority was shown for overall mechanical performance, the locking plate groups exhibited higher shear and rotational stability than the neutralization plate.

Keywords: Ankle; Locking plate; Neutralization plate; Osteoporosis; Synthetic bone.

MeSH terms

  • Ankle Fractures* / diagnostic imaging
  • Ankle Fractures* / surgery
  • Biomechanical Phenomena
  • Bone Plates
  • Fibula / surgery
  • Fracture Fixation, Internal / methods
  • Humans
  • Osteoporotic Fractures* / surgery