The objective was to assess the instrumental validity and the test-retest reliability of a low-cost hand-held push dynamometer adapted from a load-cell based hanging scale (tHHD) to collect compressive forces in different ranges of compressive forces. Three independent raters applied 50 pre-established compressions each on the tHHD centered on a force platform in three distinct ranges: ~70 N, ~160 N, ~250 N. Knee isometric strength was also assessed on 19 subjects in two sessions (48 h apart) using the tHHD anchored by an inelastic adjustable strap. Knee extension and flexion were assessed with the participant seated on a chair with the feet resting on the floor, knees, and hips flexed at 90°. The isometric force peaks were recorded and compared. The ICC and the Cronbach's α showed excellent consistency and agreement for both instrumental validity and test-retest reliability (range: 0.89-0.99), as the correlation and determination coefficients (range: 0.80-0.99). The SEM and the MDC analysis returned adequate low values with a coefficient of variation less than 5%. The Bland-Altman results showed consistency and high levels of agreement. The tHHD is a valid method to assess the knee isometric strength, showing portability, cost-effectiveness, and user-friendly interface to provide an effective form to assess the knee isometric strength.
Keywords: isometric contraction; knee assessment; muscle strength.