Abstract

Introduction: Fully hydroxyapatite-coated titanium alloy double-tapered stems have been successful implant designs for THA. The Novation Element Stem (Exactech, Gainesville, FL) is one of several of these implants following a modified Corail design philosophy. Although a great deal is known about the results of the Corail stem, little is known about the more recent Corail-inspired implants. The authors evaluated the clinical outcomes, radiographic findings and survival statistics of the collarless version of the Element Stem when used routinely in a diverse patient population.

Methods: A retrospective review was performed for all primary THR cases from 2010 to 2018. Patient demographics and HHS/OHS/patient satisfaction scores were obtained. Radiographs were evaluated for stem subsidence, radiolucencies, and bone responses to the implant.

Results: One hundred and fifty seven Element Stems were implanted with greater than 2 years of radiographic follow-up, while 105 Element Stems implanted during the study period had a survival of 5 years or more. Average HHS was 91.7 and average OHS was 44.0. 54 of 157 hips had radiolucencies, all of which included a proximal zonal radiolucency. Average subsidence was 3.33 mm; 18 hips had subsidence > 4 mm. There were 8 revisions for: aseptic loosening (4), trauma-related peri-prosthetic fractures, and early proximal-medial fractures. At 5 years, the all-cause survival rate is 92.4%, 96.2% based only on aseptic loosening and 98.1% based only on proximal medial fractures.

Conclusion: Clinical outcomes using the collarless Novation Element Stem are good, but early proximal medial fracture is still a factor in patients with poor-quality bone. Proximal radiolucencies progressing to aseptic loosening post-operatively are also a concern. The 11.5% rate of subsidence is comparable to other fully HA-coated collarless stems. This study provides a thorough critical analysis of outcomes and midterm survival data of this dual tapered-wedge fully HA-coated collarless stem used routinely in a diverse patient population.

Keywords: Aseptic loosening; Cementless collarless femoral stem; HA coating; Mid-term outcomes; Outcomes; Press fit femoral stem.

MeSH terms

  • Arthroplasty, Replacement, Hip* / methods
  • Femur / surgery
  • Follow-Up Studies
  • Fractures, Bone* / surgery
  • Hip Prosthesis*
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Treatment Outcome