When local impedance meets contact force: preliminary experience from the CHARISMA registry

J Interv Card Electrophysiol. 2022 Apr;63(3):749-758. doi: 10.1007/s10840-022-01163-7. Epub 2022 Mar 24.

Abstract

Purpose: Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have emerged as a viable real-time indicator of tissue characteristics and the consequent durability of the lesions created. We investigated the impact of catheter-tissue contact force (CF) on LI behavior during pulmonary vein isolation (PVI).

Methods: Forty-five consecutive patients of the CHARISMA registry undergoing de novo AF radiofrequency (RF) catheter ablation with a novel open-irrigated-tip catheter endowed with CF and LI measurement capabilities (Stablepoint™ catheter, Boston Scientific) were included.

Results: A total of 2895 point-by-point RF applications were analyzed (RF delivery time (DT) = 8.7±4s, CF = 13 ±±8 g, LI drop = 23 ±±7 Ω). All PVs were successfully isolated in an overall procedure time of 118 ±±34 min (fluoroscopy time = 13 ±±8 min). The magnitude of LI drop weakly correlated with CF (r = 0.13, 95% confidence interval (CI): 0.09 to 0.16, p < 0.0001), whereas both CF and LI drop inversely correlated with DT (r = -0.26, 95%CI: -0.29 to -0.22, p < 0.0001 for CF; r = -0.36, 95%CI: -0.39 to -0.33, p < 0.0001 for LI). For each 10 g of CF, LI drop markedly increased from 22.4 ± 7 Ω to 24.0 ± 8 Ω at 5 to 25 g CF intervals (5-14 g of CF vs 15-24 g of CF, p < 0.0001), whereas it showed smooth transition over 25 g (24.8 ± 7Ω at ≥ 25 g CF intervals, p = 0.0606 vs 15-24 g of CF). No major complications occurred during the procedures or within 30 days.

Conclusions: CF significantly affects LI drop and probable consequent lesion formation during RF PVI. The benefit of higher contact (> 25 g) between the catheter and the tissue appears to have less impact on LI drop.

Trial registration: Catheter Ablation of Arrhythmias With High Density Mapping System in the Real World Practice (CHARISMA). URL: http://clinicaltrials.gov/ Identifier: NCT03793998.

Keywords: Atrial fibrillation; Catheter ablation; Contact force; Lesion formation; Local impedance; PVI.