Valvulotomy in in situ vein bypasses performed by angioscope-assisted laser probe

J Surg Res. 1987 Apr;42(4):440-5. doi: 10.1016/0022-4804(87)90180-6.

Abstract

The objective of this study was to evaluate the use of an angioscope-assisted laser hot-tip probe for performing valvulotomies in in situ nonreversed canine veins used for arterial bypasses. In Group 1, 9 valvulotomies were performed in 7 bypass veins with a 2-mm probe activated by 12 W power, resulting in 4 vein perforations, 1 vein thrombosis, and 4 in situ grafts patent at removal after 2-28 days. In Group 2, 10 valvulotomies in 9 bypass veins were performed with a 1.5-mm probe at 14 W with no vein perforations, 3 vein thromboses, and 6 in situ grafts patent at removal after an interval of 1 to 6 weeks. We conclude that valvulotomy within in situ vein bypasses can be successfully performed using a 1.5-mm laser hot-tip probe at 14 W power, with patency of the grafts demonstrated at 6 weeks. Angioscope-assisted laser valvulotomy using a laser probe may significantly reduce the length of incisions and the time required for in situ vein bypasses and improve the technical accuracy of the procedure.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Blood Vessel Prosthesis
  • Dogs
  • Femoral Artery / surgery*
  • Femoral Vein / surgery*
  • Femoral Vein / transplantation
  • Laser Therapy / adverse effects
  • Laser Therapy / methods*
  • Rupture
  • Thrombosis / etiology
  • Vascular Patency
  • Vascular Surgical Procedures / instrumentation
  • Vascular Surgical Procedures / methods*