Microporous affinity membranes and their incorporation into microfluidic devices for monitoring of therapeutic antibodies

Talanta. 2023 Jan 15:252:123842. doi: 10.1016/j.talanta.2022.123842. Epub 2022 Aug 17.

Abstract

Control of monoclonal antibody (mAb) concentrations in serum is important for maintaining the safety and efficacy of these lifesaving therapeutics. Point-of-care (POC) quantification of therapeutic mAbs could ensure that patients have effective mAb levels without compromising safety. This work uses mimotope-functionalized microporous alumina affinity membranes in vertical flow assays for detection and quantitation of therapeutic mAbs. Selective capture of bevacizumab from 1000:1 diluted serum or plasma and binding of a fluorescently labelled anti-human IgG secondary antibody enable fluorescence-based analysis of bevacizumab at its therapeutically relevant concentration range of ∼50-300 μg/mL. The assay results in a linear relationship between the fluorescence intensity of the antibody capture spot and the bevacizumab concentration. A simple prototype microfluidic device containing these membranes allows washing, reagent additions and visualization of signal within 15 min using a total of 5 mL of fluid. The prototype devices can monitor physiologically relevant bevacizumab levels in diluted serum, and future refinements might lead to a POC device for therapeutic drug monitoring.

Keywords: Microfluidics; Mimotopes; Monoclonal antibodies (mAbs); Point-of-care; Therapeutic drug monitoring; Vertical flow assays.