Peri-procedural and mid-term follow-up age-related differences in leadless pacemaker implantation: Insights from a multicenter European registry

Int J Cardiol. 2023 Jan 15:371:197-203. doi: 10.1016/j.ijcard.2022.09.026. Epub 2022 Sep 15.

Abstract

Background: Age-related differences on leadless pacemaker (LP) are poorly described. Aim of this study was to compare clinical indications, periprocedural and mid-term device-associated outcomes in a large real-world cohort of LP patients, stratified by age at implantation.

Methods: Two cohorts of younger and older patients (≤50 and > 50 years old) were retrieved from the iLEAPER registry. The primary outcome was to compare the underlying indication why a LP was preferred over a transvenous PM across the two cohorts. Rates of peri-procedural and mid-term follow-up major complications as well as LP electrical performance were deemed secondary outcomes.

Results: 1154 patients were enrolled, with younger patients representing 6.2% of the entire cohort. Infective and vascular concerns were the most frequent characteristics that led to a LP implantation in the older cohort (45.8% vs 67.7%, p < 0.001; 4.2% vs 16.4%, p = 0.006), while patient preference was the leading cause to choose a LP in the younger (47.2% vs 5.6%, p < 0.001). Median overall procedural (52 [40-70] vs 50 [40-65] mins) and fluoroscopy time were similar in both groups. 4.3% of patients experienced periprocedural complications, without differences among groups. Threshold values were higher in the younger, both at discharge and at last follow-up (0.63 [0.5-0.9] vs 0.5 [0.38-0-7] V, p = 0.004).

Conclusion: When considering LP indications, patient preference was more common in younger, while infective and vascular concerns were more frequent in the older cohort. Rates of device-related complications did not differ significantly. Younger patients tended to have a slightly higher pacing threshold at mid-term follow-up.

Keywords: Age differences; Complications, pacing threshold; Leadless pacemaker; Young patients.

Publication types

  • Multicenter Study

MeSH terms

  • Cardiac Pacing, Artificial / adverse effects
  • Equipment Design
  • Humans
  • Middle Aged
  • Pacemaker, Artificial* / adverse effects
  • Time Factors
  • Treatment Outcome