Background: Transvenous lead extraction (TLE) has evolved as one of the most crucial treatment options for patients with cardiac device-related systemic infection (CDRSI).

Objective: The aim of this study was to characterize the procedural outcome and risk factors of patients with CDRSI undergoing TLE.

Methods: A subgroup analysis of patients with CDRSI of the GALLERY (GermAn Laser Lead Extraction RegistrY) database was performed. Predictors for complications, procedural failure, and all-cause mortality were evaluated.

Results: A total of 722 patients (28.6%) in the GALLERY had "systemic infection" as extraction indication. Patients with CDRSI were older (70.1 ± 12.2 years vs 67.3 ± 14.3 years; P < .001) and had more comorbidities than patients with local infections or noninfectious extraction indications. There were no differences in complete procedural success (90.6% vs 91.7%; P = .328) or major complications (2.5% vs 1.9%; P = .416) but increased procedure-related (1.4% vs 0.3%; P = .003) and all-cause in-hospital mortality (11.1% vs 0.6%; P < .001) for patients with CDRSI. Multivariate analyses revealed lead age ≥10 years as a predictor for procedural complications (odds ratio [OR] 3.23; 95% confidence interval [CI] 1.58-6.60; P = .001). Lead age ≥10 years (OR 2.57; 95% CI 1.03-6.46; P = .04) was also a predictor for procedural failure. We identified left ventricular ejection fraction <30% (OR 1.70; 95% CI 1.00-2.99; P = .049), age ≥75 years (OR 2.1; 95% CI 1.27-3.48; P = .004), chronic kidney disease (OR 1.92; 95% CI 1.17-3.14; P = .01), and overall procedural complications (OR 5.15; 95% CI 2.44-10.84; P < .001) as predictors for all-cause mortality.

Conclusion: Patients with CDRSI undergoing TLE demonstrate an increased rate of all-cause in-hospital, as well as procedure-related mortality, despite having comparable procedural success rates. Given these data, it seems paramount to develop preventive strategies to detect and treat CDRSI in its earliest stages.

Keywords: Cardiac device–related infection; Cardiac implantable electronic device; Endocarditis; Lead management; Transvenous lead extraction.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Child
  • Comorbidity
  • Defibrillators, Implantable* / adverse effects
  • Device Removal / adverse effects
  • Heart Diseases* / etiology
  • Humans
  • Pacemaker, Artificial* / adverse effects
  • Retrospective Studies
  • Stroke Volume
  • Treatment Outcome
  • Ventricular Function, Left