Development and validation of a device to measure the force applied to the coxofemoral joint during stress radiography for early diagnosis of canine hip dysplasia

Med Eng Phys. 2023 Feb:112:103953. doi: 10.1016/j.medengphy.2023.103953. Epub 2023 Jan 9.

Abstract

Hip dysplasia is a common orthopaedic condition in dogs and stress radiography is the best diagnostic tool for early diagnosis. Objective force guidelines are lacking, leaving room for errors and fraud during screening. Our objective was to develop an accurate and validated measuring device that allows quantification of the applied force in vivo in real-time during stress radiographic imaging. A two-step approach was followed. First, four load cells were incorporated in the original Vezzoni Modified Badertscher Distension Device (VMBDD) and a dedicated computer program was developed. In vitro evaluations of the accuracy demonstrated a trueness of 0.19 N (0.1%FS) and precision of 0.26 N (0.2%FS) for the individual loadcells. The trueness and precision of the assembled VMBDmD were 0.02 N (0.02%FS) and 0.52 N (0.38%FS). Secondly, the modified device was tested on several cadavers. The device was similar in use as the VMBDD, did not interfere with radiographic acquisition, gave the operator real-time feedback, and linked the force with the radiograph. Altogether, we describe the accuracy of the VMBDmD and have evaluated its use in cadavers. We saw that the device successfully quantified and stored the applied force in real-time during stress radiography.

Keywords: Canine hip dysplasia (CHD); Force measuring device; Stress radiography.

MeSH terms

  • Animals
  • Dogs
  • Early Diagnosis
  • Hip Dysplasia, Canine* / diagnostic imaging
  • Hip Joint
  • Radiography