Inferior vena cava filter retrievals using advanced techniques: a systematic review

Diagn Interv Radiol. 2023 May 31;29(3):500-508. doi: 10.4274/dir.2022.22908. Epub 2023 Feb 27.

Abstract

This study aimed to assess the safety of complex inferior vena cava (IVC) filter retrieval techniques through a systematic review of published literature. Using PubMed, a systematic review was conducted in line with the 2020 Preferred Reporting Items for Systematic Review and Meta-Analysis guidelines to identify articles published through April 2020 that reported on complex IVC filter retrieval techniques in >5 patients. Case reports, review papers, and studies that did not report on primary outcomes or variables of interest were excluded. Risk of bias was assessed using a modified Newcastle-Ottawa Quality Assessment scale. Pooled success and complication rates were calculated for the overall number of complex retrieval attempts as well as for each filter type and each complex retrieval method. Sixteen fair-quality and three good-quality studies met the inclusion criteria, with 758 patients (428 female) who had undergone 770 advanced retrieval attempts. The mean age of the patients was 46.5 ± 7.1 years (range: 14.1-90), and the mean dwell time was 602.5 ± 388.6 days (range: 5-7336). Regarding filters, 92.6% (702/758) were retrievable and 7.4% (56/758) were permanent. Indications for complex retrieval included the failure of standard retrieval (89.2%; 676/758) and tilting or embedding in the caval wall (53.8%; 408/758); 92.6% (713/770) of the advanced retrieval attempts were successful. The pooled success rate was 92.0% (602/654) for retrievable filters and 96.4% (53/55) for permanent filters (P = 0.422). Only 2.8% (21/758) of patients experienced major complications, and the major complication rate was not significantly associated with filter type (P = 0.183). Advanced techniques for IVC filter retrieval appear safe for the retrieval of retrievable filters and certain permanent filters, with a low short-term major complication rate. Further studies on complex retrieval techniques used to remove permanent filters should be conducted to clarify their safety with respect to filter type.

Keywords: Advanced filter retrieval; complex filter retrieval; pulmonary embolism; thromboembolism; venous thromboembolism.

Publication types

  • Systematic Review

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Device Removal / methods
  • Female
  • Humans
  • Middle Aged
  • Pulmonary Embolism* / prevention & control
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome
  • Vena Cava Filters* / adverse effects
  • Vena Cava, Inferior / diagnostic imaging
  • Vena Cava, Inferior / surgery