Objectives: DNA integrity number (DIN) is a metric for assessing DNA degradation, calculated based on electrophoresis using the Agilent TapeStation System. The utility of DIN as a diagnostic indicator of sufficient DNA quality in clinical next-generation sequencing (NGS) has not been well described.
Methods: We evaluated the DINs of 166 tumor formalin-fixed, paraffin-embedded (FFPE) tissue samples submitted for 124-gene panel sequencing. We also investigated a new metric on the electropherogram that could improve the predictive accuracy of the DIN.
Results: A DIN cutoff of 2.5 discriminated samples with successful analysis (n = 143) from samples with failed analysis (n = 23), with a sensitivity of 0.84 and a specificity of 0.78 (area under the curve [AUC] = 0.88). The DIN was positively correlated with the mean coverage (r = 0.72, P < .0001) but could not discriminate success from failure when the DIN was below 2.5 (negative predictive value, 0.44). We introduced a new metric, the peak/base ratio, that distinguished success from failure with higher accuracy than the DIN (cutoff = 1.6; sensitivity = 0.98, specificity = 0.83, and AUC =0.96).
Conclusions: To predict successful NGS, the DNA quality of FFPE tissue can be easily and reliably assessed using the DIN and peak/base ratio.
Keywords: DNA degradation; DNA integrity number (DIN); electrophoresis; formalin-fixed; next-generation sequencing (NGS); paraffin-embedded tissue; peak/base ratio.