Impact of High Implantation of Transcatheter Aortic Valve on Subsequent Conduction Disturbances and Coronary Access

JACC Cardiovasc Interv. 2023 May 22;16(10):1192-1204. doi: 10.1016/j.jcin.2023.03.021.

Abstract

Background: Data regarding the impact of high transcatheter heart valve (THV) implantation on coronary access after transcatheter aortic valve replacement (TAVR) as assessed by postimplantation computed tomography (CT) are scarce.

Objectives: The authors sought to assess the impact of high THV implantation on coronary access after TAVR.

Methods: We included 160 and 258 patients treated with Evolut R/PRO/PRO+ and SAPIEN 3 THVs, respectively. In the Evolut R/PRO/PRO+ group, the target implantation depth was 1 to 3 mm using the cusp overlap view with commissural alignment technique for the high implantation technique (HIT), whereas it was 3 to 5 mm using 3-cusp coplanar view for the conventional implantation technique (CIT). In the SAPIEN 3 group, the HIT employed the radiolucent line-guided implantation, whereas the central balloon marker-guided implantation was used for the CIT. Post-TAVR CT was performed to analyze coronary accessibility.

Results: HIT reduced the incidence of new conduction disturbances after TAVR for both THVs. In the Evolut R/PRO/PRO+ group, post-TAVR CT showed that the HIT group had a higher incidence of the interference of THV skirt (22.0% vs 9.1%; P = 0.03) and a lower incidence of the interference of THV commissural posts (26.0% vs 42.7%; P = 0.04) with access to 1 or both coronary ostia compared with the CIT group. These incidences were similar between the HIT and CIT groups in the SAPIEN 3 group (THV skirt: 0.9% vs 0.7%; P = 1.00; THV commissural tabs: 15.7% vs 15.3%; P = 0.93). In both THVs, CT-identified risk of sinus sequestration in TAVR-in-TAVR was significantly higher in the HIT group compared with the CIT group (Evolut R/PRO/PRO+ group: 64.0% vs 41.8%; P = 0.009; SAPIEN 3 group: 17.6% vs 5.3%; P = 0.002).

Conclusions: High THV implantation substantially reduced conduction disturbances after TAVR. However, post-TAVR CT revealed that there is a risk for unfavorable future coronary access after TAVR and sinus sequestration in TAVR-in-TAVR. (Impact of High Implantation of Transcatheter Heart Valve during Transcatheter Aortic Valve Replacement on Future Coronary Access; UMIN000048336).

Keywords: commissural alignment; coronary access; permanent pacemaker; redo TAVR; transcatheter aortic valve replacement.

MeSH terms

  • Aortic Valve* / diagnostic imaging
  • Aortic Valve* / surgery
  • Humans
  • Tomography, X-Ray Computed
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Treatment Outcome

Associated data

  • UMIN-CTR/UMIN000048336