Use of Omniflow® II Biosynthetic Graft for the Treatment of Vascular Graft and Endograft Infections

Ann Vasc Surg. 2023 Nov:97:410-418. doi: 10.1016/j.avsg.2023.05.020. Epub 2023 May 25.

Abstract

Background: Vascular graft/endograft infection is a rare but life-threatening complication of cardiovascular surgery and remains a surgical challenge. Several different graft materials are available for the treatment of vascular graft/endograft infection, each having its own advantages and disadvantages. Biosynthetic vascular grafts have shown low reinfection rates and could be a potential second best after autologous veins in the treatment of vascular graft/endograft infection. Therefore, the aim of our study was to evaluate the efficacy and morbidity of Omniflow® II for the treatment of vascular graft/endograft infection.

Methods: A multicenter retrospective cohort study was performed to evaluate the use of Omniflow® II in the abdominal and peripheral region to treat vascular graft/endograft infection between January 2014 and December 2021. Primary outcome was recurrent vascular graft infection. Secondary outcomes included primary patency, primary assisted patency, secondary patency, all-cause mortality, and major amputation.

Results: Fifty-two patients were included with a median follow-up duration of 26.5 (10.8-54.8) months. Nine (17%) grafts were implanted in intracavitary position and 43 (83%) in peripheral position. Most grafts were used as femoral interposition (n = 12, 23%), femoro-femoral crossover (n = 10, 19%), femoro-popliteal (n = 8, 15%), and aorto-bifemoral (n = 8, 15%) graft. Fifteen (29%) grafts were implanted extra-anatomically and 37 (71%) in situ. Eight patients (15%) presented with reinfection during follow-up, most of these patients received an aorto-bifemoral graft (n = 3, 38%). Intracavitary vascular grafting had a 33% (n = 3) reinfection rate and peripheral grafting 12% (n = 5; P = 0.025). The estimated primary patencies at 1, 2, and 3 years were 75%, 72%, and 72% for peripherally located grafts and 58% (at all timepoints) for intracavitary grafts (P = 0.815). Secondary patencies at 1, 2, and 3 years were 77% (at all timepoints) for peripherally located prostheses and 75% (at all timepoints) for intracavitary prostheses (P = 0.731). A significantly higher mortality during follow-up was observed in patients who received an intracavitary graft compared to patients with a peripheral graft (P = 0.003).

Conclusions: This study highlights the efficacy and safety of the Omniflow® II biosynthetic prosthesis for the treatment of vascular graft/endograft infection, in absence of suitable venous material, with acceptable reinfection, patency, and freedom of amputation prevalences, especially in replacing peripheral vascular graft/endograft infection. However, a control group with either venous reconstruction or another alternative graft is needed to make firmer conclusions.

Publication types

  • Multicenter Study

MeSH terms

  • Blood Vessel Prosthesis / adverse effects
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Humans
  • Prosthesis-Related Infections* / surgery
  • Reinfection
  • Retrospective Studies
  • Treatment Outcome
  • Vascular Diseases* / surgery
  • Vascular Patency