IN.PACT AV Access Randomized Trial of Drug-Coated Balloons for Dysfunctional Arteriovenous Fistulae: Clinical Outcomes through 36 Months

J Vasc Interv Radiol. 2023 Dec;34(12):2093-2102.e7. doi: 10.1016/j.jvir.2023.07.007. Epub 2023 Jul 17.

Abstract

Purpose: To present the 36-month outcomes of the prospective randomized IN.PACT AV Access study of participants with obstructive de novo or restenotic native upper extremity arteriovenous dialysis fistula lesions treated with drug-coated balloon (DCBs) or standard percutaneous transluminal angioplasty (PTA) following successful high-pressure PTA.

Materials and methods: Participants at 29 international sites were randomized 1:1 to receive an IN.PACT AV DCB (n = 170) or undergo PTA (n = 160). The outcomes through 36 months included target lesion primary patency (TLPP) and access circuit primary patency (ACPP) (composites of clinically driven target lesion or access circuit revascularization and/or access circuit thrombosis), number of reinterventions, and serious adverse events involving the access circuit.

Results: TLPP was 52.1% in the DCB group compared with 36.7% in the PTA group through 24 months and 43.1% in the DCB group compared with 28.6% in the PTA group through 36 months (both log-rank P < .001). ACPP was 39.4% in the DCB group compared with 25.3% in the PTA group through 24 months and 26.4% in the DCB group compared with 16.6% in the PTA group through 36 months (both log-rank P < .001). Cumulative incidence of access circuit thrombosis through 36 months was 8.2% in the DCB group compared with 18.3% in the PTA group (log-rank P = .040). Cumulative incidence of mortality through 36 months was 26.6% in the DCB group compared with 30.8% in the PTA group (log-rank P = .71).

Conclusions: This study demonstrated superior TLPP and ACPP with DCBs compared with PTA, with no difference in mortality through 3 years. Access circuit thrombosis was statistically significantly higher in the PTA group at 3 years.

Publication types

  • Randomized Controlled Trial
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Angioplasty, Balloon* / adverse effects
  • Coated Materials, Biocompatible
  • Femoral Artery
  • Humans
  • Peripheral Arterial Disease* / diagnostic imaging
  • Peripheral Arterial Disease* / therapy
  • Popliteal Artery
  • Prospective Studies
  • Single-Blind Method
  • Thrombosis* / diagnostic imaging
  • Thrombosis* / etiology
  • Thrombosis* / therapy
  • Time Factors
  • Treatment Outcome
  • Vascular Access Devices*
  • Vascular Patency

Substances

  • Coated Materials, Biocompatible