Background: The American Academy of Oral and Maxillofacial Radiology established an ad hoc committee to draft evidence-based recommendations and clinical guidance for the application of patient contact shielding during dentomaxillofacial imaging.
Types of studies reviewed: The committee reviewed monographs and reports from radiation protection organizations and studies that reported radiation dose to gonads, breasts, and thyroid gland from dentomaxillofacial imaging.
Results: Considering the absence of radiation-induced heritable effects in humans and the negligible dose to the gonads and fetus from dentomaxillofacial imaging, the committee recommends discontinuing shielding of the gonads, pelvic structures, and fetuses during all dentomaxillofacial radiographic imaging procedures. On the basis of radiation doses from contemporaneous maxillofacial imaging, the committee considered that the risks from thyroid cancer are negligible and recommends that thyroid shielding not be used during intraoral, panoramic, cephalometric, and cone-beam computed tomographic imaging.
Practical implications: This position statement informs and educates the reader on evolving radiation protection practices and provides simple, unequivocal guidance to dental personnel to implement these guidelines. State and local authorities should be contacted to update regulations to reflect these recommendations.
Keywords: Radiation effects; lead apron; radiation protection; radiation shielding; thyroid collar.