Background Mechanistic insight into the high failure rate of TOF-Cuff® (RGB Medical Devices, Madrid, Spain) measurements on the lower leg is unclear. Aims We aimed to determine whether materials applied to pseudo-skin can reduce the impedance between a model arm and TOF-Cuff® electrodes and whether a material between TOF-Cuff® electrodes and the patient's skin surface decreases the skin-TOF-Cuff® electrode impedance within the appropriate range. Methods This was a combination of an in vitro study using non-living materials and a prospective observational clinical study. Eight patients aged > 70 years who had undergone elective surgery were eligible. One of the primary outcomes was whether water, electrocardiogram (ECG) cream, or ECG gel applied on the pseudo-skin could reduce the impedance between the model arm and the TOF-Cuff® electrodes in the in vitro study. Another was whether a material between the TOF-Cuff® electrodes and the patient's skin surface decreased the skin-TOF-Cuff® electrode impedance to an appropriate level of less than 5,000 Ω in the clinical study. Results The application of water, ECG cream, and ECG gel similarly reduced the impedance values within the electrical circuit in the in vitro study. ECG cream application between the patient's skin surface and the TOF-Cuff® electrodes decreased the skin-TOF-Cuff® electrode impedance (median (interquartile range (IQR)) Ω) from 8,600 (6,450 to 9,775) to 2,000 (1,600 to 2,600) (P = 0.012) in surgical patients. Conclusion ECG cream application between the patient's skin surface and the TOF-Cuff® electrodes decreased the skin-TOF-Cuff® electrode impedance appropriately, and thus, the application can facilitate precise TOF-Cuff® measurements in patients.
Keywords: electrocardiogram cream; impedance; muscle relaxant; neuromuscular block monitoring; neuromuscular monitoring.
Copyright © 2023, Sugimura et al.