Development of a Magnetoresistive-Based Wearable Eye-Tracking System for Oculomotor Assessment in Neurological and Otoneurological Research-Preliminary In Vivo Tests

Brain Sci. 2023 Oct 10;13(10):1439. doi: 10.3390/brainsci13101439.

Abstract

Over the past 20 years, several eye-tracking technologies have been developed. This article aims to present a new type of eye tracker capable of producing detailed information on eye and head movements using an array of magnetoresistive detectors fixed on the patient's head and a small magnet inserted into a contact lens, adapted to the curvature of the cornea of the subject. The software used for data analysis can combine or compare eye and head movements and can represent them as 2D or 3D images. Preliminary data involve an initial patient who was asked to perform several tasks to establish the accuracy, reliability, and tolerance of the magnetic eye tracker and software. The tasks included assessment of saccadic eye movements and pursuit, "drawing" alphabetic shapes or letters, and reading. Finally, a Head Impulse Test (HIT) was performed to estimate the VOR gain, comparing the standard deviation established via vHIT with that established via this magnetic eye tracker (mHIT). This prototypical device is minimally invasive, lightweight, relatively cheap, and tolerable, with a high degree of reliability and precision. All these characteristics could lead to the future use of the magnetic eye tracker in neurological and otoneurological fields.

Keywords: eye movements; eye tracker; search coil; vHIT.

Grants and funding

This research received no external funding.