ExacTrac X-Ray 6D Imaging During Stereotactic Body Radiation Therapy of Spinal and Nonspinal Metastases

Technol Cancer Res Treat. 2023 Jan-Dec:22:15330338231210786. doi: 10.1177/15330338231210786.

Abstract

The objective was to investigate the possibility of using ExacTrac X-ray (ETX) for 6D image guidance in stereotactic body radiation therapy (SBRT) of bone metastasis and to propose a patient management protocol. The analyses were first obtained from measurements on a pelvic phantom and on 19 patients treated for bone metastasis. The phantom study consisted of applying known offsets and evaluating the ETX level of accuracy, where results were compared with kV-cone beam computed tomography (kV-CBCT). Two groups of patients, 10 spinal and 9 nonspinal SBRT cases, were analyzed to evaluate ETX imaging for different bone localisations. A comparison was made between kV-CBCT and ETX prior to the treatment fractions. During treatments, two other kV-CBCT/ETX image pairs were also acquired and a total of 224 shifts were compared. A second study, using the ETX monitoring module, analyzed the intrafraction motion of 8 other patients. In the phantom study, the root mean square (RMS) of the translational and rotational discrepancies between ETX and kV-CBCT were < 0.6 mm and < 0.4°, respectively. For both groups of patients, the RMS of the discrepancies observed between the two imaging systems were greater than the phantom experiment while still remaining < 1 mm and < 0.7°. In the nonspinal group, three patients (2 scapulas and 1 humerus) did not have consistent shift values with ETX due to a lack of anatomical information. When ETX monitoring was used during irradiation, the setup errors measured were on average less than 1 mm/1°. The results obtained validated the use of ETX for 6D image guidance during bone SBRT. Real-time tracking of the target position improves the accuracy of the irradiation. This strategy allowed for faster correction of out-of-tolerance positioning errors. The registration of bone lesions with poor anatomical information is a limitation of this 2D-kV imaging system.

Keywords: ExacTrac X-ray 6D; IGRT; bone SBRT; intra-fraction motion; kV-CBCT.

MeSH terms

  • Cone-Beam Computed Tomography / methods
  • Humans
  • Phantoms, Imaging
  • Radiography
  • Radiosurgery* / methods
  • Radiotherapy Planning, Computer-Assisted / methods
  • X-Rays