The present study investigated the cyclic fatigue properties of titanium alloys (Ti-6Al-4V and Ti-6Al-7Nb) as implant materials and compared their properties with those of commercially pure titanium. Ti-6Al-4V and Ti-6Al-7Nb cylinders with diameters of 3.0 mm were examined. The surfaces of the cylinders were roughened by sand blasting with alumina particles and acid etching. Static and cyclic tests were performed according to ISO 14801:2016. The yield force in the static test (YS) was measured in 5 specimens of each alloy using a universal testing machine. The yield force in a cyclic test (YC) was measured in 20 specimens of each alloy using the staircase method, which involved applying a cyclic load at a frequency of 10 Hz for 106 cycles. After the cyclic loading tests, cross-sections of the specimens were examined under an optical microscope. The YS values for Ti-6Al-4V and Ti-6Al-7Nb were 1463 N±93 N and 1405 N±79 N, respectively, and the YC values were 870 N±58 N and 853 N±202 N, respectively. Microscopic observation revealed cracks on the tensile side of some of the specimens, including run outs and failures. The results of this study suggest that the YC values for Ti-6Al-4V and Ti-6Al-7Nb were 40% less than those for YS. The yield force of Grade-4 Cp-Ti significantly decreased after cyclic loading. The YC values for Ti-6Al-4V and Ti-6Al-7Nb were approximately 900 N, which was markedly greater than that for commercially pure, Grade-4 titanium (700 N).
Keywords: Cyclic fatigue; Implants; Titanium alloys.