Ventricular assist device using a thoracotomy-based implant technique: Multi-Center Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy (HM3 SWIFT)

J Thorac Cardiovasc Surg. 2024 Nov;168(5):1474-1484.e12. doi: 10.1016/j.jtcvs.2024.02.013. Epub 2024 Feb 15.

Abstract

Objectives: The HeartMate 3 (Abbott) left ventricular assist device provides substantial improvement in long-term morbidity and mortality in patients with advanced heart failure. The Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy study compares thoracotomy-based implantation clinical outcomes with standard median sternotomy.

Methods: We conducted a prospective, multicenter, single-arm study in patients eligible for HeartMate 3 implantation with thoracotomy-based surgical technique (bilateral thoracotomy or partial upper sternotomy with left thoracotomy). The composite primary end point was survival free of disabling stroke (modified Rankin score >3), or reoperation to remove or replace a malfunctioning device, or conversion to median sternotomy at 6-months postimplant (elective transplants were treated as a success). The primary end point (noninferiority, -15% margin) was assessed with >90% power compared with a propensity score-matched cohort (ratio 1:2) derived from the Multi-Center Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3 continued access protocol.

Results: The study enrolled 102 patients between December 2020 and July 2022 in the thoracotomy-based arm at 23 North American centers. Follow-up concluded in December 2022. In the Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy study group, noninferiority criteria was met (absolute between-group difference, -1.2%; Farrington Manning lower 1-sided 95% CI, -9.3%; P < .0025) and event-free survival was not different (85.0% vs 86.2%; hazard ratio, 1.01; 95% CI, 0.58-2.10). Length of stay with thoracotomy-based implant was longer (median, 20 vs 17 days; P = .03). No differences were observed for blood product utilization, adverse events (including right heart failure), functional status, and quality of life between cohorts.

Conclusions: Thoracotomy-based implantation of the HeartMate 3 left ventricular assist device is noninferior to implantation via standard full sternotomy. This study supports thoracotomy-based implantation as an additional standard for surgical implantation of the HeartMate 3 left ventricular assist device.

Keywords: HeartMate 3; LVAD; minimally invasive; outcomes; thoracotomy.

Publication types

  • Multicenter Study
  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Female
  • Heart Failure* / mortality
  • Heart Failure* / physiopathology
  • Heart Failure* / surgery
  • Heart-Assist Devices*
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design
  • Prosthesis Implantation* / adverse effects
  • Prosthesis Implantation* / instrumentation
  • Prosthesis Implantation* / methods
  • Prosthesis Implantation* / mortality
  • Sternotomy* / adverse effects
  • Thoracotomy* / adverse effects
  • Time Factors
  • Treatment Outcome
  • United States
  • Ventricular Function, Left