Noninvasive measurement of ambulatory venous pressure via column interruption duration in chronic venous disease

J Vasc Surg Venous Lymphat Disord. 2024 Jul;12(4):101861. doi: 10.1016/j.jvsv.2024.101861. Epub 2024 Feb 28.

Abstract

Background: Column interruption duration (CID) is a noninvasive surrogate for venous refill time (VFT), a parameter used in ambulatory venous pressure measurement. CID is more accurate than invasive VFT measurement because it avoids errors involved with indirect access of the deep system through the dorsal foot vein. The aim of this retrospective single center study is to analyze the clinical usefulness of CID in assessment of chronic venous disease (CVD).

Methods: A total of 1551 limbs (777 patients) were referred with CVD symptoms over a 5-year period (2018-2023); CID, air plethysmography, and duplex reflux data were analyzed. Of these limbs, 679 had supine venous pressure data as well. The pathology was categorized as obstruction if supine peripheral venous pressure was >11 mm Hg and as reflux if duplex reflux time in superficial or deep veins was >1 second. CID was measured via Doppler monitoring of flow in the great saphenous vein (GSV) and one of the paired posterior tibial (PT) veins near the ankle in the erect posture. The calf is emptied by rapid inflation cuff. CID is the time interval in seconds when cephalad venous flow in great saphenous vein and posterior tibial veins reappear after calf ejection. A CID <20 seconds in either vein is abnormal similar to the threshold used in VFT measurement.

Results: Thirty-two percent of the limbs had obstruction, 17% had reflux, and 37% had a combination; 14% had neither. Higher clinical-etiology-anatomy-pathophysiology (CEAP) clinical classes (C4-6) were prevalent in 44% of pure reflux, significantly less (P < .0001) than in pure obstruction (73%) or obstruction plus reflux subsets (72%), partly reflecting distribution of pathology. There is a progressive increase in supine venous pressure and abnormal CID (P < .0001 and P < .0001, respectively) in successive CEAP clinical class. No such correlation between CEAP and any of the reflux severity grading methods (reflux segment score, Venous Filling Index, and Kistner axial grading) was observed. Abnormal CID (55%) was more prevalent in higher CEAP classes (>4) (P < .0001) than in lesser clinical classes (0-2) or limbs with neither obstruction nor reflux (P < .01).

Conclusions: Obstruction seems to be a more dominant pathology in clinical progression among CEAP clinical classes than reflux. CID is abnormal in both obstructive and refluxive pathologies and may represent a common end pathway for similar clinical manifestations (eg, ulcer). These data suggest a useful role for CID measurement in clinical assessment of limbs with CVD.

Keywords: Ambulatory venous pressure; CID; Column interruption duration; Venous obstruction; Venous pressure; Venous reflux.

MeSH terms

  • Adult
  • Aged
  • Blood Pressure Monitoring, Ambulatory
  • Chronic Disease
  • Female
  • Humans
  • Male
  • Middle Aged
  • Plethysmography*
  • Predictive Value of Tests
  • Reproducibility of Results
  • Retrospective Studies
  • Saphenous Vein / diagnostic imaging
  • Saphenous Vein / physiopathology
  • Time Factors
  • Ultrasonography, Doppler, Duplex
  • Venous Insufficiency* / diagnostic imaging
  • Venous Insufficiency* / physiopathology
  • Venous Pressure