Evaluation of Dynamic Multi-Leaf Collimator (MLC) versus Fixed MLC for Intensity Modulated Radiotherapy (IMRT) Using the Agility 160-Leaf Collimator

Asian Pac J Cancer Prev. 2024 Jul 1;25(7):2467-2474. doi: 10.31557/APJCP.2024.25.7.2467.

Abstract

Aim: This study aimed to evaluate the efficacy of static or step-and-shoot intensity-modulated radiotherapy (ssIMRT) and dynamic intensity-modulated radiotherapy (dIMRT) delivery techniques for various treatment sites.

Materials and methods: The treatment planning system (TPS) was utilized to develop optimal treatment plans for twenty-seven patients selected for this comparative study, including nine with head and neck cancer, nine with prostate cancer, and nine with cervical cancer. The prescribed doses were 7000cGy/33fr, 7425cGy/33fr, and 5000cGy/25fr for the nasopharynx, prostate, and cervix cases, respectively, in both ssIMRT and dIMRT delivery techniques. Plans were generated using the Monaco treatment planning system with a 6MV photon beam and nine equidistant fields. Plan evaluation criteria included dose-volume histogram analysis, dose homogeneity index, conformity index, radiation delivery time, and monitor unit requirements.

Results: All plans were optimized to ensure that 98% of the planning target volume (PTV) received at least 95% of the prescribed dose, while meeting the planning objectives for organs at risk. dIMRT plans exhibited superior conformity (CI = 0.85 ± 0.05) compared to ssIMRT plans (CI = 0.79 ± 0.08), with statistically significant differences (P < 0.01). Inhomogeneity within the PTV was significantly higher in ssIMRT plans (HI = 0.10 ± 0.02) compared to dIMRT plans (HI = 0.09 ± 0.01), with a significant difference (P < 0.01). Delivery time per fraction was significantly lower in dIMRT compared to ssIMRT (P < 0.01). Furthermore, dIMRT plans required a higher mean monitor unit value (1335.4 ± 172.2) compared to ssIMRT plans (974.4 ± 133.6) with a significant difference (P < 0.001).

Conclusion: The findings of this study indicate that dIMRT provides improved target coverage, homogeneity, and conformity while reducing treatment delivery time compared to ssIMRT.

Keywords: Conformity Index; Delivery Time; Homogeneity Index; Monitor Unit.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Female
  • Follow-Up Studies
  • Head and Neck Neoplasms* / radiotherapy
  • Humans
  • Male
  • Organs at Risk* / radiation effects
  • Prognosis
  • Prostatic Neoplasms* / radiotherapy
  • Radiotherapy Dosage*
  • Radiotherapy Planning, Computer-Assisted* / methods
  • Radiotherapy, Intensity-Modulated* / methods
  • Uterine Cervical Neoplasms / radiotherapy