In Intensive Care Unit (ICU), the settings of the critical alarms should be sensitive and patient-specific to detect signs of deteriorating health without ringing continuously, but alarm thresholds are not always calibrated to operate this way. An assessment of the connection between critical alarm threshold settings and the patient-specific variables in ICU would deepen our understanding of the issue. The aim of this retrospective descriptive and exploratory study was to assess this relationship using a large cohort of ICU patient stays. A retrospective study was conducted on some 70,000 ICU stays taken from the MIMIC-IV database. Critical alarm threshold values and threshold modification frequencies were examined. The link between these alarm threshold settings and 30 patient variables was then explored by computing the Shapley values of a Random Tree Forest model, fitted with patient variables and alarm settings. The study included 57,667 ICU patient stays. Alarm threshold values and alarm threshold modification frequencies exhibited the same trend: they were influenced by the vital sign monitored, but almost never by the patient's overall health status. This exploratory study also placed patients' vital signs as the most important variables, far ahead of medication. In conclusion, alarm settings were rigid and mechanical and were rarely adapted to the evolution of the patient. The management of alarms in ICU appears to be imperfect, and a different approach could result in better patient care and improved quality of life at work for staff.
Keywords: Clinical practices; Critical alarms; Intensive care unit.
© 2024. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.