Objectives: To develop and evaluate a deep learning (DL) model to reduce metal artefacts originating from the exomass in cone-beam computed tomography (CBCT) of the jaws.
Methods: Five porcine mandibles, each featuring six tubes filled with a radiopaque solution, were scanned using four CBCT units before and after the incremental insertion of up to three titanium, titanium-zirconium, and zirconia dental implants in the exomass of a small field of view. A conditional denoising diffusion probabilistic model, using DL techniques, was employed to correct images from exomass-related metal artefacts across the CBCT units and implant scenarios. Three examiners independently scored the image quality of all datasets, including those without an implant (ground truth), with implants in the exomass (original), and DL-generated ones. Quantitative analysis compared contrast-to-noise ratio (CNR) to validate artefact reduction using repeated measures analysis of variance in a factorial design followed by Tukey test (α = 0.05).
Results: The visualisation of the hard tissues and overall image quality was reduced in the original and increased in the DL-generated images. The score variation observed in the original images was not observed in the DL-generated images, which generally scored higher than the original images. DL-generated images revealed significantly greater CNR than both the ground truth and their corresponding original images, regardless of the material and quantity of dental implants and the CBCT unit (p < 0.05). Original images revealed significantly lower CNR than the ground truth (p < 0.05).
Conclusions: The developed DL model demonstrated promising performance in correcting exomass-related metal artefacts in CBCT of the jaws.
Keywords: artefacts; artificial intelligence; cone-beam computed tomography; deep learning; dental implants.
© The Author(s) 2024. Published by Oxford University Press on behalf of the British Institute of Radiology and the International Association of Dentomaxillofacial Radiology.