To further improve analytical accuracy in clinical chemistry, proficiency testing needs amelioration in the quality of materials tested and in target value assignment. To obtain information on the actual state-of-the-art in the Lombardy region of Italy, and to examine the behavior of different types of control materials (fresh-frozen human sera and lyophilized materials), we developed the following experimental design. Two human serum pools and two lyophilized sera were distributed to 32 laboratories for determination of glucose, creatinine, cholesterol, sodium, potassium, and gamma-glutamyltransferase (gamma-GT). Each analyte was measured in triplicate on each of 3 days. Target values for the controls were obtained with Reference Methods. The results show a good intralaboratory precision for every component but some accuracy problems for glucose, creatinine, cholesterol, and gamma-GT. Lyophilized materials showed some commutability problems for glucose, electrolytes, and gamma-GT, mainly with dry chemistry technology.