Continuous venovenous hemodiafiltration in infants and children

Am J Kidney Dis. 1995 Jan;25(1):17-21. doi: 10.1016/0272-6386(95)90618-5.

Abstract

Continuous venovenous hemodiafiltration (CVVHD) is not commonly used in pediatric intensive care units due to the lack of suitable equipment needed for this technique of renal replacement therapy (RRT). We have used an adapted hemodialysis machine that includes a blood pump controller, an air leak detector, and a venous pressure monitor over the past year in the pediatric intensive care unit. Blood lines available for hemodialysis were used for CVVHD, limiting the extracorporeal circuit volume to 38 mL, which allows for CVVHD capability in an infant as small as 4.5 kg without a blood-primed circuit. We have compared this experience to previous continuous arteriovenous hemodiafiltration (CAVHD) at our institution. The two groups (CVVHD and CAVHD) were similar in age, weight, blood pressure, and indication for RRT. There was significantly less number of hemofilters used, an improved number of hours per hemofilter, and a significantly less change of RRT modality due to ineffective dialysis (CVVHD 0% v CAVHD 32%) when using CVVHD. Furthermore, an average of 48% less heparin was used in the CVVHD population. We conclude that CVVHD can be safely and effectively carried out in infants and small children with less heparinization, no need for arterial access, and less risk of ineffective RRT.

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Child
  • Child, Preschool
  • Female
  • Hemodiafiltration / instrumentation
  • Hemodiafiltration / methods*
  • Humans
  • Infant
  • Male
  • Retrospective Studies
  • Treatment Outcome