Effectiveness of noninvasive programmed stimulation for initiating ventricular tachyarrhythmias in patients with third-generation implantable cardioverter defibrillators

Pacing Clin Electrophysiol. 1994 Sep;17(9):1462-8. doi: 10.1111/j.1540-8159.1994.tb01510.x.

Abstract

Previous generations of implantable cardioverter defibrillators (ICDs) required invasive electrophysiological testing to assess defibrillator function. Newer third-generation ICDs include the capability for performing noninvasive programmed stimulation (NIPS) and may reduce the need for invasive studies to assess tachycardia recognition and antitachycardia therapy algorithms. The effectiveness of ICD-based NIPS for the induction of ventricular arrhythmias has not, however, been formally assessed. Third-generation ICDs were implanted in 79 patients, who underwent a total of 166 postoperative defibrillator tests. NIPS with rapid ventricular pacing was performed in all patients in an attempt to induce ventricular fibrillation. In patients with prior sustained uniform ventricular tachycardia, programmed stimulation with up to three extrastimuli was performed in order to attempt to initiate the clinical ventricular tachcardia. Ventricular fibrillation was induced with NIPS in 146 of 166 studies (88%). Ventricular tachycardia was initiated with NIPS in 104 of 123 studies (85%). The type of defibrillator and the use of endocardial or epicardial rate sensing/pacing leads did not influence the efficacy of NIPS. NIPS with third-generation ICDs is generally effective at inducing ventricular fibrillation and clinically relevant ventricular tachycardias, and reduces the need to perform invasive electrophysiological testing following device implantation. In a minority of patients temporary transvenous pacing catheters must still be used to facilitate arrhythmia induction.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Cardiac Pacing, Artificial / methods*
  • Defibrillators, Implantable*
  • Electric Stimulation
  • Electrodes, Implanted
  • Equipment Design
  • Female
  • Humans
  • Male
  • Middle Aged
  • Postoperative Care
  • Tachycardia, Ventricular / physiopathology*
  • Tachycardia, Ventricular / therapy*
  • Ventricular Fibrillation / physiopathology
  • Ventricular Fibrillation / therapy