A comparison of defibrillation thresholds was made using biphasic and monophasic shocks delivered by a nonthoracotomy lead system in 2 clinically distinct groups of patients. The first group were patients receiving an implantable cardioverter-defibrillator who were studied before surgery with their chests closed. The second group were patients undergoing coronary artery bypass grafting (CABG) who were studied before surgery with their chests open but reapproximated. Biphasic defibrillation thresholds (stored energy) were significantly (p < 0.001) less than monophasic ones in subjects with the implantable cardioverter-defibrillator (12.3 +/- 5.3 vs 21.1 +/- 9.3 J) or CABG (14.6 +/- 7.1 vs 24.2 +/- 12.6 J). These values are less than were previously reported with a similar nonthoracotomy lead configuration. There were no significant differences between the 2 groups in all measurements derived from corresponding shock waveforms, although impedance tended to be greater in patients with CABG. However, subjects with CABG had greater left ventricular ejection fractions and did not have history of potentially lethal ventricular arrhythmias. Despite these differences, the conclusion that biphasic shocks are more effective would have been made in a study of either group alone. It is concluded that patients with CABG who have not had preceding potentially lethal ventricular arrhythmias may be a potential source of surrogate subjects for defibrillation research such as epicardial mapping, which requires that the chest be open.