Wedge factors: dependence on depth and field size

Radiother Oncol. 1996 Apr;39(1):31-4. doi: 10.1016/0167-8140(95)01714-3.

Abstract

Wedge factor (WF) variation with field size and depth has been analysed for 6 and 15 MV X-ray beams. The measured field size effect is 1.4% in the worst case, whereas the WF shows a linear dependence with depth giving a maximum variation of 0.22%/cm (60 degrees wedge, 6 MV beam). An empirical correction for WF depth dependence on wedge thickness is proposed.

MeSH terms

  • Absorption
  • Algorithms
  • Humans
  • Phantoms, Imaging
  • Radiotherapy Dosage*
  • Radiotherapy, High-Energy*
  • Scattering, Radiation