Microdosimetric evaluation of relative biological effectiveness for 103Pd, 125I, 241Am, and 192Ir brachytherapy sources

Int J Radiat Oncol Biol Phys. 1996 Oct 1;36(3):689-97. doi: 10.1016/s0360-3016(96)00374-4.

Abstract

Purpose: To determine the microdosimetric-derived relative biological effectiveness (RBE) of 103Pd, 125I, 241Am, and 192Ir brachytherapy sources at low doses and/or low dose rates.

Methods and materials: The Theory of Dual Radiation Action can be used to predict expected RBE values based on the spatial distribution of energy deposition at microscopic levels from these sources. Single-event lineal energy spectra for these isotopes have been obtained both experimentally and theoretically. A grid-defined wall-less proportional counter was used to measure the lineal energy distributions. Unlike conventional Rossi proportional counters, the counter used in these measurements has a conducting nylon fiber as the central collecting anode and has no metal parts. Thus, the Z-dependence of the photoelectric effect is eliminated as a source of measurement error. Single-event spectra for these brachytherapy sources have been also calculated by: (a) the Monte Carlo code MCNP to generate the electron slowing down spectrum, (b) transport of monoenergetic electron tracks, event by event, with our Monte Carlo code DELTA, (c) using the concept of associated volume to obtain the lineal energy distribution f(y) for each monoenergetic electron, and (d) obtaining the composite lineal energy spectrum for a given brachytherapy source based on the electron spectrum calculated at step (a).

Results: Relative to 60Co, the RBE values obtained from this study are: 2.3 for 103Pd, 2.1 for 125I, 2.1 for 241Am, and 1.3 for 192Ir.

Conclusions: These values are consistent with available data from in vitro cell survival experiments. We suggest that, at least for these brachytherapy sources, microdosimetry may be used as a credible alternative to time-consuming (and often uncertain) radiobiological experiments to obtain information on radiation quality and make reliable predictions of RBE in low dose rate brachytherapy.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Americium*
  • Brachytherapy
  • Half-Life
  • Iodine Radioisotopes*
  • Iridium Radioisotopes*
  • Palladium*
  • Radioisotopes*
  • Relative Biological Effectiveness*