Magnetization transfer imaging in vivo of the rat brain at 4.7 T: interpretation using a binary spin-bath model with a superLorentzian lineshape

Magn Reson Med. 1997 Dec;38(6):974-80. doi: 10.1002/mrm.1910380617.

Abstract

Proton magnetization transfer contrast (MTC) imaging, using continuous wave off-resonance irradiation, was performed on the rat brain in vivo at 4.7 Tesla. The observed MTC was studied in three different brain regions: the corpus callosum, the basal ganglia, and the temporal lobe. By systematically varying the offset frequency and the amplitude of the RF irradiation, the observed signal intensities for each region of interest were modeled using a system including free water and a pool of protons with restricted motions (R. M. Henkelman, X. Huang, Q. Xiang, G. J. Stanisz, SD Swanson, M. J. Bronskill, Magn. Res. Med. 29, 759 (1993)). Most of the relaxation parameters of both proton pools remained fairly constant for the three regions of interest, with a T2 value of about 9 micros for the immobilized protons, whereas the rate of exchange increased significantly from the temporal lobe to the corpus callosum. The optimal acquisition parameters for the improved MTC under steady-state saturation were found to be 2-10 kHz offset frequency and 500-800 Hz RF irradiation amplitude. Conversely, an irradiation amplitude of 3 kHz at an offset frequency of 12 kHz is required to minimize the direct effect of off-resonance irradiation. Such an approach could be extended to human brain imaging with the aim of characterizing tissue-specific disease.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain / anatomy & histology*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Models, Theoretical
  • Rats
  • Rats, Wistar