[Acetabulum deformations after implantation of a cemented cup with or without metal-back component. An in vitro comparative study of monopodal load]

Rev Chir Orthop Reparatrice Appar Mot. 1997;83(5):409-15.
[Article in French]

Abstract

Purpose of the study: The purpose of this study was to find a biomechanical explanation for a clinical failure of metal backed acetabular components. Periacetabular deformations were measured on fresh cadaver bones equipped with strain-gauge rosettes.

Material and methods: Two skeletons, including pelvic bone, the two last lumbar vertebrae, and both femurs were maintained in unipodal equilibrium using metallic cables for muscle simulation. Loads were applied up to 700 N the approximating average body weight. A comparative study of periacetabular deformations was performed: right sides of each pelvic bone were implanted with conventional cemented implants, and left sides with metal backed cemented implants of same diameter and size.

Results: Significant differences were noted: whereas no modification was brought by implantation of conventional cemented sockets, periacetabular deformations were reduced and smoothed after implantation of cemented metal backed implants. Such results are in agreement with recently published data obtained, using the finite element analysis method.

Discussion: A study of pelvic ring and acetabular walls displacement was performed, but the authors couldn't find any clinical relevance to this biomechanical study.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Acetabulum* / physiopathology
  • Arthroplasty, Replacement, Hip*
  • Biomechanical Phenomena
  • Bone Cements / therapeutic use
  • Cadaver
  • Humans
  • Middle Aged
  • Models, Anatomic
  • Prosthesis Failure
  • Stress, Mechanical
  • Weight-Bearing

Substances

  • Bone Cements