We describe an analytical system for the rapid identification of proteins by correlation of tandem mass spectra with protein sequence databases. The system consists of an integrated solid phase microextraction/capillary zone electrophoresis peptide separation device that is connected through a microelectrospray ion source to a tandem mass spectrometer. The limits of detection are 660 amol of sample at a concentration limit of < 33 amol/microliters for peptide mass measurement, and < 10 fmol of sample, at a concentration limit of < 300 amol/microliters for peptide analysis by collision-induced dissociation. Using this system, we have identified low nanogram amounts of yeast proteins separated by high-resolution two-dimensional gel electrophoresis.